Sensorex Sensorex Differential Amplifier, p/n# EM801 Sensorex Differential Battery-Powered pH/ORP Amplifier for use with S8000 Series, 8 Pin DIN Connecto.. Product #: EM801 Regular price: $217.75 $217.75 In Stock

Sensorex Differential Amplifier, p/n# EM801

$217.75

Product Code: EM801
Brand: Sensorex
Availability: In Stock

Sensorex Differential Battery-Powered pH/ORP Amplifier for use with S8000 Series. Ryton PPS Body, 8 Pin DIN Connector, 0 to 80 C Operating Temp, Diameter 36.83 mm, Length 100.58 mm.


The battery cannot be replaced. This module must be paired with electrode adapter EA899TC (has solution ground input, necessary for correct function of the differential amplifier circuit). If the module fails (reads 7pH always or 0mV for ORP), then replace with a new one. Typical life is about three years from the date stamped on the module (MMYY).


Features:


Testing Parameter pH or ORP (REDOX)
Operating Temp. Range 0 Degrees to 80 Degrees C
Body Material Ryton PPS
Diameter 36.83 mm
Length 100.58 mm
Connector 8 Pin DIN
For Use With S8000 Platform
Shelf Life 12 mo


Applications:


  Industrial


 Wastewater


 Dirty Applications



- +